Tecfluid FLOMID-MX Manual do Utilizador Página 8

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 20
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 7
8
4 SET-UP
Set-up mode can be accessed by pressing the
and push buttons at the same time, to
avoid unwanted changes due to accidental access.
4.1 Flow rate set-up
F l o w r a t e s e t u p
In this submenu the parameters which affect the flow rate calculations are introduced:
Sensor factor
Electronics factor
Measuring units
Number of decimal points
Nominal flow rate
Minimum flow rate (dropout)
Pressing the
push button, the screen for entering data of the first parameter appears.
4.1.1 Sensor factor
To start with we must enter the factor to adapt the control unit to the sensor to be used.
In the first screen the calibration factor which appears on the sensor identification label should
be introduced.
S e n s o r f a c t o r
> 1 . 00000
When the screen appears the cursor of the digit to be modified is the underline of the digit. The
push button is used to increment the value of the digit. Once the desired value is reached, by
pressing the
push button the cursor will pass to the next digit to be modified. On the last digit by
pressing the
push button the cursor will pass to the first digit. On pressing the push button the
data displayed on the screen will be stored in memory and the next programming screen will appear.
If you want to exit the programming screen without storing the data in memory, even if changes
have been made, then press the
push button.
NOTE: A factor Fc = 0.00000 will not be accepted.
4.1.2 Electronics Factor
Once the sensor factor has been stored the next screen for introducing the electronics calibration factor
will appear. In this screen the calibration factor on the identification label of the electronics unit must be
introduced.
E l e c t . f a c t o r
> 1 . 00000
Vista de página 7
1 2 3 4 5 6 7 8 9 10 11 12 13 ... 19 20

Comentários a estes Manuais

Sem comentários